Magnetic circular dichroism (MCD) is the differential absorption of left and right circularly polarized light induced by a longitudinal magnetic field between different electronic states.

SETUP
Similar to CD regarding using the PEM.


A SIMPLE EXAMPLE FOR MEASURING  THE ZEEMAN EFFECT
Electronic transition for an atom: 1S -> 1P
(one of the possibilities from s2 -> sp without spin-orbit coupling)



The three states of 1P will be split in the presence of a longitudinal magnetic field as shown above. The transition select rules govern that left and right circularly polarized light be absorbed at different energies. The resulted MCD spectrum is a derivative-like band whose amplitude is proportional to the Zeeman splitting.

The Zeeman splitting can be measured from MCD without resolving the hyperfine peaks. In the above figures, the Zeeman splitting is greatly exaggerated to illustrate the physical content. In reality, the MCD signal for diamagnetic species is generally on the order of 10-3 to 10-6. Furthermore, MCD can be observed in the IR region for vibrational transitions where it is called MVCD.

TYPICAL PEMs 
I/FS50 with Magnetic Field Compatible Head Option (MFC)
II/FS42 with Magnetic Field Compatible Head Option (MFC)

APPLICATIONS
In physics, chemistry and biochemistry; hyperfine structures of atoms and molecules; analytical applications in biochemistry.
   RESOURCES

   PEM Newsletter #2:  Simultaneous MCD & Double Absorption Spectra   



FURTHER READINGS
P. J. Stephens, Adv. Chem. Phys. 35, 197 (1976).

S. B. Piepho and P. N. Schatz– Group Theory in Spectroscopy with Applications to Magnetic Circular Dichroism, (Wiley-Interscience, New York, 1983).

T. A. Keiderling, J. Chem. Phys. 75, 3639 (1981).

B. Wang, C. N. Tam and T. A. Keiderling, Phys. Rev. Lett. 71, 979 (1993).